Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8021B: (2 year) CAL VER /5700 INSTRUMENT: Fluke 8022B: (2 year) CAL VER /5700 DATE: 10-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 29 NUMBER OF LINES: 113 CONFIGURATION: Fluke 5700A ============================================================================= # # Source: # Fluke 8021B instruction manual (PN 616037 July 1981). # Fluke 8022B instruction manual (PN 616045 July 1981). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 57_8021.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 MMFC C2 1.004 HEAD EQUIPMENT SETUP 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Ambient temperature: 21C - 25C. 1.006 DISP Relative humidity: less than 80%. 1.006 DISP Allow the UUT to stabilize. 1.007 DISP Turn the UUT on. 1.008 HEAD {RESISTANCE TESTS} 1.009 PIC 57_8021 1.010 HEAD RESISTANCE TESTS: {2 kOhm Range} 1.011 5700 2 0.000kZ +0.001U CW 2.001 HEAD RESISTANCE TESTS: {200 Ohm Range} 2.002 5700 200 0.0Z +0.2U CW 3.001 5700 200 190.0Z 0.9U CW 4.001 HEAD RESISTANCE TESTS: {2 kOhm Range} 4.002 5700 2 1.900kZ 0.005U CW 5.001 HEAD RESISTANCE TESTS: {20 kOhm Range} 5.002 5700 20 19.00kZ 0.05U CW 6.001 HEAD RESISTANCE TESTS: {200 kOhm Range} 6.002 5700 200 190.0kZ 0.5U 2W 7.001 HEAD RESISTANCE TESTS: {2000 kOhm Range} 7.002 5700 2000 1900kZ 39U 2W 8.001 HEAD RESISTANCE TESTS: {20 MOhm Range} 8.002 5700 20 10.00MZ 0.2U 2W 9.001 JMPT 11.001 Fluke 8022B: (2 year) CAL VER /5700 9.002 HEAD {CONTINUITY TEST AUDIBLE TONE} 9.003 RNG 2kZ 9.004 DISP Depress the AC/DC switch to activate the audible tone. 9.005 5700 0Z S CW 9.006 EVAL Is there an {audible tone}? 10.001 5700 110V 60H S 2W 10.002 EVAL Is UUT {beeping at a rate of approximately 5 to 10 per sec}. 11.001 HEAD {DC VOLTAGE TESTS} 11.002 HEAD DC VOLTAGE TESTS: {200mV Range} 11.003 5700 200 190.0mV 0.6U 2W 12.001 5700 200 -190.0mV 0.6U 2W 13.001 HEAD DC VOLTAGE TESTS: {2V Range} 13.002 5700 2 0.000V 0.001U 2W 14.001 5700 2 1.900V 0.006U 2W 15.001 HEAD DC VOLTAGE TESTS: {20V Range} 15.002 5700 20 19.00V 0.06U 2W 16.001 HEAD DC VOLTAGE TESTS: {200V Range} 16.002 5700 200 190.0V 0.6U 2W 17.001 HEAD DC VOLTAGE TESTS: {1000V Range} 17.002 5700 1000 1000V 3U 2W 18.001 HEAD {AC VOLTAGE TESTS} 18.002 HEAD AC VOLTAGE TESTS: {200mV Range} 18.003 RNG 200mV AC 18.004 5700 0.0mV S 2W 18.005 MEMI Enter UUT reading in millivolts AC: 18.006 MEME 18.007 MEMC mV +0.2U 19.001 5700 200 190.0mV 1.6U 100H 2W 20.001 5700 200 19.0mV 0.3U 100H O 2W 21.001 HEAD AC VOLTAGE TESTS: {2V Range} 21.002 5700 2 1.900V 0.016U 100H 2W 22.001 HEAD AC VOLTAGE TESTS: {20V Range} 22.002 5700 20 19.00V 0.16U 100H 2W 23.001 HEAD AC VOLTAGE TESTS: {200V Range} 23.002 5700 200 190.0V 1.6U 100H 2W 24.001 HEAD AC VOLTAGE TESTS: {750V Range} 24.002 5700 750 750V 9U 400H 2W 25.001 HEAD {DC CURRENT TESTS} 25.002 HEAD DC CURRENT TESTS: {2mA Range} 25.003 5700 2 1.900mA 0.015U 2W 26.001 HEAD DC CURRENT TESTS: {20mA Range} 26.002 5700 20 -19.00mA 0.15U 2W 27.001 HEAD DC CURRENT TESTS: {200mA Range} 27.002 5700 200 190.0mA 1.5U 2W 28.001 HEAD DC CURRENT TESTS: {2000mA Range} 28.002 5700 2000 1900mA 15U 2W 29.001 END